| Physical specifications |
| Dimensions |
70 x 85 x 140cm |
| Weight |
65kg |
| Casters |
Front wheels with lock |
| Screen |
LCD |
| Display |
10.1 inch |
| Display graphics |
Waveforms of P-T, F-T, V-T; Loops of P-V, V-F, F-P |
| Working specifications |
| Power |
AC 110V~240V, 50Hz / 60Hz |
| Battery Power |
≥ 180 mins |
| Pipeline supply |
| Gas configuration |
O₂, N₂O, Air |
| Pipeline input range |
280-600kPa (40~85psi) |
| Ventilator Specifications |
| Applications |
| Patient Range |
Adult / Paediatric |
| Modes of ventilation |
| VCV |
Volume Control Ventilation |
| SIMV |
Synchronized Intermittent Mandatory Ventilation |
| MANUAL |
Manual |
| STAND-BY |
STAND-BY |
| A/C |
Assist / Control |
| Ventilator parameter |
| Working type |
Electronic control, pneumatic driven |
| Tidal Volume range |
20-1500ml |
| Pressure range |
5-60 cm H₂O |
| Respiration Rate Range |
1-100bpm (1-40bpm under SIMV) |
| I: E Range |
4:1-1:8 |
| Ptrigger Range |
-20-0cm H₂O |
| Minute volume range |
>18L/Min |
| SIGH range |
0-5/100 |
| System Standard |
| Hypoxic guard system |
N₂O cut-off valve, O₂ concentration > 21 % |
| Safety valve |
<12.5kPa |
| O₂ Flush |
25-75L/min |
| Flowmeter |
4 tubes, cascade 0-15L/min O₂, 0-15L/min N₂O, 0-15L/min Air (Optional) |
| Volume of CO₂ absorber |
2 L |
| Vaporizer |
1 Selectetac, maximum 2 mounts (Isoflurane/Enflurane /Sevoflurane /Halothane) |
| Monitoring |
VT, MV, Pmax, Pmean, Lung compliance, Freq, ETCO₂, FiO₂ (Optional) |
| Alarm Limits |
| Audio/Visual |
No tidal volume, MV, Paw |
|
Air failure, AC power failure, Battery low |
| Pre-setting functions |
| Language |
ENGLISH |
| Calibration |
Automatic |
Reviews